Central Venous Pressure From Peripherally Inserted Central catheters Correlates Well With That of Centrally Inserted Central Catheters FREE TO VIEW

Mariam Alansari, MD, FRCSI; Mohammed Hijazi, MD, FCCP*
Author and Funding Information

KFSH RC, Riyadh, Saudi Arabia


Chest. 2004;126(4_MeetingAbstracts):873S. doi:10.1378/chest.126.4_MeetingAbstracts.873S-a
Text Size: A A A
Published online


PURPOSE:  To compare CVP measured by CICCs through the distal port of a triple lumen catheter size; with the transducer at the level of the heart; with CVP measured via distal port of the PICCs with transducer level at –4, -2, 0, +2, and +4 from heart.

METHODS:  Measurements of central venous pressure were recorded from patients who had an indwelling CICC and PICC concomitantly. Pressure measurements were taken from 18-gauge dual-lumen PICCs and from 16-gauge CICCs all with continuous pressure infusion devices. Central venous pressure measurements, using a pressure infusion device to overcome the natural resistance of the PICC, were taken from both the CICC and the PICC at the level of the heart simultaneously. Two more readings were taken from the PICC, both at 2 and 4 cm above and below the heart respectively. This cycle was repeated 6 times in each patient. Measurements were recorded at end-expiration.

RESULTS:  120 pairs of CVP readings were obtained from 20 paients. Among these, 12 had femoral access and 8 were mixed between internal jugular and subclavian.For measuring and comparing the agreement between CVP from CICC and CVP FROM PICC, concordence correlation (Rc) was calculated. The Rc was 0.99 at the level of the heart (Rc value between 0.9-1, means substantial agreement). See the tableTable 1

Rc values in the 20 patients

MeasurementsRCPeripherally Inserted Central Catheter at 4 cm above the heart level0.906Peripherally Inserted Central Catheter at 2 cm above the heart level0.972Peripherally Inserted Central Catheter0.992Peripherally Inserted Central Catheter at 2 cm below the heart level0.941Peripherally Inserted Central Catheter at 4 cm below the heart level0.834below.

CONCLUSION:  Central venous pressure measured through a peripherally inserted central catheter at the level of the heart had substantial agreement with central venous pressure measured through centrally inserted central catheter, at the same level. This applies to femoral internal jagular and subclavian access.

CLINICAL IMPLICATIONS:  Measuring CVP from PICCs is possible without subjecting the patient to invasive procedures.With an additional benefit of central venous pressure monitoring from PICCs, more institutions may adopt increasing use of PICCs in acute settings. Nurse led PICC line insertion was found to be successful, which would be of great help in decreasing the work load on physician in intensive care units. Table 2

Rc values in the subgroup of patients (femoral acess vs neck lines)

MeasurementsFemoral linesIJ and SC lines*Peripherally Inserted Central Catheter at 4 cm above the heart level0.89610.9239Peripherally Inserted Central Catheter at 2 cm above the heart level0.96730.9822Peripherally Inserted Central Catheter0.99220.9935Peripherally Inserted Central Catheter at 2 cm below the heart level0.93930.9515Peripherally Inserted Central Catheter at 4 cm below the heart level0.83260.8442

IJ = intrnal jugular SC = SubClavian

DISCLOSURE:  M. Hijazi, None.

Wednesday, October 27, 2004

12:30 PM- 2:00 PM




Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

CHEST Journal Articles
PubMed Articles
  • CHEST Journal
    Print ISSN: 0012-3692
    Online ISSN: 1931-3543