Looking at Transpulmonary Thermodilution Curves : The Cross-Talk Phenomenon FREE TO VIEW

Frédéric Michard, MD, PhD
Author and Funding Information

Massachusetts General Hospital-Harvard Medical School Boston, MA

Correspondence to: Frédéric Michard, MD, PhD, Department of Anesthesia and Critical Care, Massachusetts General Hospital-Harvard Medical School, 55 Fruit St, Boston, MA 02114; e-mail: fmichard@partners.org



Chest. 2004;126(2):656-657. doi:10.1378/chest.126.2.656-a
Text Size: A A A
Published online

To the Editor:

Transpulmonary thermodilution has been shown to be an easy and reliable technique for hemodynamic monitoring in critically ill patients and is being used increasingly.14 After injection of a cold saline solution bolus via a central venous catheter, a thermistor-tipped arterial catheter (usually femoral) is connected to a specific monitor (PiCCOplus; Pulsion Medical Systems; Munich, Germany), which records the downstream temperature changes, ie, a transpulmonary thermodilution curve. The mathematical analysis of the curve allows the computation of cardiac output, heart blood volume (global end-diastolic volume), and an estimation of extravascular lung water.,14

Recently, the mere observation of the transpulmonary thermodilution curve has been shown to be useful to detect and monitor right-to-left intracardiac shunting.5 Indeed, in case of right-to-left intracardiac shunt, one part of the cold indicator passes through the atrial septum and rapidly reaches the arterial thermistor. As a result, the thermodilution curve appears prematurely and becomes biphasic with two humps (“camel” curve).5

I would like to report such a camel thermodilution curve that was not related to a right-to-left intracardiac shunt. This curve was recorded in a patient with septic shock instrumented with a right femoral thermistor-tipped arterial catheter and a right femoral venous catheter. Both catheters were of the same length (20 cm). When the cold saline solution was injected through the femoral venous line, the transpulmonary thermodilution curve was highly suggestive of right-to-left intracardiac shunting (Fig 1 ). However, the patient was not hypoxemic and contrast echocardiography as well as color Doppler imaging did not reveal any intracardiac shunt. Because both arterial and venous femoral catheters were on the same side and of the same length, it was hypothesized that the high concentration of cold indicator at the site of injection (the tip of the venous catheter) may induce significant temperature changes in the femoral artery, ie, at the level of the arterial catheter tip equipped with a thermistor (cross-talk phenomenon). To confirm this hypothesis, central venous injections were also performed through an internal jugular line. In this case, the shape of the transpulmonary thermodilution curve was normal (Fig 1).

Looking at transpulmonary thermodilution curves can be very useful to diagnose intracardiac shunts.5 However, a cross-talk phenomenon may result in a double-hump thermodilution curve wrongly suggestive of right-to-left intracardiac shunting. Therefore, the use of venous and thermistor-tipped arterial catheters on the same side and of the same length should be avoided in patients monitored with transpulmonary thermodilution.

Figure Jump LinkFigure 1. Transpulmonary thermodilution curves recorded after the injection of a cold saline solution bolus in the 20-cm femoral venous catheter (F) and in the jugular venous catheter (J). The double-hump curve is not explained by a right-to-left intracardiac shunt but by a cross-talk phenomenon: the decrease in blood temperature during the cold bolus injection is directly transmitted to the thermistor-tipped arterial catheter.Grahic Jump Location
Michard, F, Perel, A (2003) Management of circulatory and respiratory failure using less invasive hemodynamic monitoring. Vincent, JL eds.Yearbook of intensive care and emergency medicine,508-520 Springer. Berlin, Germany:
 
Sakka, SG, Reinhart, K, Meier-Hellmann, A Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients.Intensive Care Med1999;25,843-846
 
Michard, F, Alaya, S, Zarka, V, et al Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock.Chest2003;124,1900-1908
 
Sakka, SG, Ruhl, CC, Pfeiffer, UJ, et al Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution.Intensive Care Med2000;26,180-187
 
Michard, F, Alaya, S, Medkour, F Monitoring right to left intracardiac shunt in acute respiratory distress syndrome.Crit Care Med2004;32,308-309
 

Figures

Figure Jump LinkFigure 1. Transpulmonary thermodilution curves recorded after the injection of a cold saline solution bolus in the 20-cm femoral venous catheter (F) and in the jugular venous catheter (J). The double-hump curve is not explained by a right-to-left intracardiac shunt but by a cross-talk phenomenon: the decrease in blood temperature during the cold bolus injection is directly transmitted to the thermistor-tipped arterial catheter.Grahic Jump Location

Tables

References

Michard, F, Perel, A (2003) Management of circulatory and respiratory failure using less invasive hemodynamic monitoring. Vincent, JL eds.Yearbook of intensive care and emergency medicine,508-520 Springer. Berlin, Germany:
 
Sakka, SG, Reinhart, K, Meier-Hellmann, A Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients.Intensive Care Med1999;25,843-846
 
Michard, F, Alaya, S, Zarka, V, et al Global end-diastolic volume as an indicator of cardiac preload in patients with septic shock.Chest2003;124,1900-1908
 
Sakka, SG, Ruhl, CC, Pfeiffer, UJ, et al Assessment of cardiac preload and extravascular lung water by single transpulmonary thermodilution.Intensive Care Med2000;26,180-187
 
Michard, F, Alaya, S, Medkour, F Monitoring right to left intracardiac shunt in acute respiratory distress syndrome.Crit Care Med2004;32,308-309
 
NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

  • CHEST Journal
    Print ISSN: 0012-3692
    Online ISSN: 1931-3543