Inter-instrument comparison of blood gas analyzers and assessment of tonometry using fresh heparinized whole human blood. FREE TO VIEW

A L Van Kessel; J H Eichhorn; J L Clausen; M E Stone; H H Rotman; R O Crapo
Chest. 1987;92(3):418-422. doi:10.1378/chest.92.3.418
Text Size: A A A
Published online

Abstract

Because of limitations associated with commercial blood gas quality controls and tonometry of stored blood, fresh heparinized blood was used to compare PO2 and PCO2 performance of ten blood gas analyzers. Function of nine gas mixer/tonometer systems was evaluated. These were used to create blood samples with target values for PO2 and PCO2. All ten analyzers had high precision; this magnified small differences between observed results and target values. Grand mean results from all ten analyzers were within 0.8 mm Hg of the target for PO2 of 40.0 and 100.0 and for PCO2 of 20.0 and 70.0 mm Hg. Eight automated blood gas analyzers gave clinically indistinguishable results for PO2 and PCO2 that were sufficiently accurate for clinical purposes. Also reported is the practicality of using tonometry of fresh heparinized blood in a per shift quality control program for PO2 and PCO2.


Figures

Tables

References

NOTE:
Citing articles are presented as examples only. In non-demo SCM6 implementation, integration with CrossRef’s "Cited By" API will populate this tab (http://www.crossref.org/citedby.html).

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging & repositioning the boxes below.

Find Similar Articles
CHEST Journal Articles
PubMed Articles
  • CHEST Journal
    Print ISSN: 0012-3692
    Online ISSN: 1931-3543